2014-09-30 06:20:14 Cryo Capability Details 10-200 K
2014-09-30 06:19:58 Cryo Capability LHe cryostat; continuous flow Oxford CF1208 continuous flow liquid helium cryostat
2014-09-30 12:41:21 Backup Services Local: CD writer (PC); Network: FTP, autobackup to server; Backup server: 4 mm DAT, 8 mm, TK50, CD writer (OpenVMS)
2014-09-30 12:41:16 Hardware DECAlpha 500/266 (OpenVMS); PC (Windows NT); Terminal server (Windows 2000); OpenVMS server access
2014-09-30 12:08:23 Software: Classification Control, data acquisition, analysis
2014-09-30 12:07:40 Lab Facilities Sample prep lab; coldroom Sample preparation lab; coldroom
2014-09-30 12:07:22 Cryo Capability LHe, continuous flow on request LHe cryostat; continuous flow
2014-09-30 12:04:42 Detector: Type Ionization Chambers, Lytle, PIPS and Vortex single element, Si-drift detector. Ionization Chambers, Lytle, PIPS and Vortex single element, Si-drift detectors
2014-09-30 12:04:14 Flux Exponent 13 ph/s at 9 keV 13 ph/s at 9 keV; for (Si(111); 4x10^12 ph/s for (Si(311)
2014-09-30 12:03:46 Mirrors M0 mirror: Bent flat single-crystal Si, Rh coated 10-18.2 keV cutoff, vertically collimating M1 mirror: Bent cylindrical single-crystal Si, Rh coated, standard focus M0 mirror: Bent flat single-crystal Si, Rh coated 10-18.2 keV cutoff, vertically collimating; M1 mirror: Bent cylindrical single-crystal Si, Rh coated, standard focus
2014-09-30 12:03:23 Crystal Type Si(111), phi=0 and Si(311) phi=0, Liq N2 cooled Si(111), phi=0 and Si(311) phi=0
2014-09-30 12:03:02 Monochromator In-vacuum; non-fixed exit slit; double flat crystal Liquid nitrogen-cooled double crystal, non-fixed exit slit
2014-09-30 12:01:38 Experiments Details Beamline 6-2 (middle hutch) is a wiggler end-station beamline dedicated for High Resolution Hard X-ray Spectroscopy. The end station combines three mutlicrystal Johann spectrometers that enable X-ray Emission Spectroscopy (XES), Resonant Inelastic X-ray Scattering (RIXS), High-Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy (HERFD-XAS) and X-ray Raman Spectroscopy (XRS) techniques. The ambient pressure operation of the end-station provides a great flexibility of implementing various sample environments (in-situ, cryogenic, He-atmosphere, furnaces, liquids, high pressurized cells, etc). A dedicated liquid-He cryostat allows for routine low temperature measurements. End station is operating with a focused x-ray beam of ~0.15x0.4 (VxH) mm2 and a flux of 2x10^13 (4x1012) when the Si111 (Si311) monochromator is used. Beam line 6-2 (middle hutch) is a wiggler end-station beam line for High Resolution Hard X-ray Spectroscopy. The end-station combines three multi-crystal analyzer spectrometers that enable X-ray Emission Spectroscopy (XES), Resonant Inelastic X-ray Scattering (RIXS), High-Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy (HERFD-XAS) and X-ray Raman Spectroscopy (XRS) techniques. An energy-dispersive von Hamos geometry spectrometer is available for time-resolved studies. The end-station provides high flexibility of implementing various sample environments, including a liquid-He cryostat for routine low temperature measurements.
2014-09-30 11:59:33 Experiments XAFS XAFS
2014-09-30 11:59:20 Status operational Operational
2014-09-30 11:59:13 Status Details open Open
2014-09-30 11:59:06 %Time Available (general use) 67 50
2014-09-30 11:58:47 Experiments Details Beamline 6-2b is a wiggler end-station beamline dedicated for High Resolution Hard X-ray Spectroscopy. The end station combines three mutlicrystal Johann spectrometers that enable X-ray Emission Spectroscopy (XES), Resonant Inelastic X-ray Scattering (RIXS), High-Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy (HERFD-XAS) and X-ray Raman Spectroscopy (XRS) techniques. The ambient pressure operation of the end-station provides a great flexibility of implementing various sample environments (in-situ, cryogenic, He-atmosphere, furnaces, liquids, high pressurized cells, etc). A dedicated liquid-He cryostat allows for routine low temperature measurements. End station is operating with a focused x-ray beam of ~0.15x0.4 (VxH) mm2 and a flux of 2x10^13 (4x1012) when the Si111 (Si311) monochromator is used. Beamline 6-2 (middle hutch) is a wiggler end-station beamline dedicated for High Resolution Hard X-ray Spectroscopy. The end station combines three mutlicrystal Johann spectrometers that enable X-ray Emission Spectroscopy (XES), Resonant Inelastic X-ray Scattering (RIXS), High-Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy (HERFD-XAS) and X-ray Raman Spectroscopy (XRS) techniques. The ambient pressure operation of the end-station provides a great flexibility of implementing various sample environments (in-situ, cryogenic, He-atmosphere, furnaces, liquids, high pressurized cells, etc). A dedicated liquid-He cryostat allows for routine low temperature measurements. End station is operating with a focused x-ray beam of ~0.15x0.4 (VxH) mm2 and a flux of 2x10^13 (4x1012) when the Si111 (Si311) monochromator is used.
2014-09-30 11:57:40 Secondary Contact: Name Dimosthenes Sokaras Dimosthenis Sokaras
2014-09-25 07:01:05 Experiments Details Beamline 6-2b is a wiggler end-station beamline dedicated for High Resolution Hard X-ray Spectroscopy. The end station combines three mutlicrystal Johann spectrometers that enable X-ray Emission Spectroscopy (XES), Resonant Inelastic X-ray Scattering (RIXS), High-Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy (HERFD-XAS) and X-ray Raman Spectroscopy (XRS) techniques. The ambient pressure operation of the end-station provides a great flexibility of implementing various sample environments (in-situ, cryogenic, He-atmosphere, furnaces, liquids, high pressurized cells, etc). A dedicated liquid-He cryostat allows for routine low temperature measurements. End station is operating with a focused x-ray beam of ~0.15x0.4 (VxH) mm2 and a flux of 2x10^13 (4x1012) when the Si111 (Si311) monochromator is used.
2014-09-25 07:00:43 Special Capabilities Beamline 6-2b is a wiggler end-station beamline dedicated for High Resolution Hard X-ray Spectroscopy. The end station combines three mutlicrystal Johann spectrometers that enable X-ray Emission Spectroscopy (XES), Resonant Inelastic X-ray Scattering (RIXS), High-Energy Resolution Fluorescence Detected X-ray Absorption Spectroscopy (HERFD-XAS) and X-ray Raman Spectroscopy (XRS) techniques. The ambient pressure operation of the end-station provides a great flexibility of implementing various sample environments (in-situ, cryogenic, He-atmosphere, furnaces, liquids, high pressurized cells, etc). A dedicated liquid-He cryostat allows for routine low temperature measurements. End station is operating with a focused x-ray beam of ~0.15x0.4 (VxH) mm2 and a flux of 2x1013 (4x1012) when the Si111 (Si311) monochromator is used. null
2014-09-25 06:59:25 Flux Exponent At 9 keV, 2 x 10^13 ph/s 13 ph/s at 9 keV
2014-09-25 06:59:13 Flux Coefficient 2
2014-09-25 06:58:52 Flux Coefficient 5
2014-09-25 06:58:08 Experiments XAFS XAFS
2014-09-25 06:57:56 Secondary Contact: Email dsokaras@slac.stanford.edu
2014-09-25 06:57:37 Secondary Contact: Name Dimosthenes Sokaras
2014-09-25 06:56:40 Software: Name ICS (instrumentation control); SPEC ICS (instrumentation control); SPEC, PyMCA, Matlab
2014-09-25 06:56:25 Software: Name ICS (instrumentation control); XAS Collect (acquisition), Ge13Plus (detector setup), EXAFSPAK (analysis) ICS (instrumentation control); SPEC
2014-09-25 06:55:59 Detector: Type Lytle and PIPS detector Ionization Chambers, Lytle, PIPS and Vortex single element, Si-drift detector.
2014-09-25 06:55:13 Goniometer none
2014-09-25 06:55:01 Energy Range (keV): Max 18.200 18.2
2014-09-25 06:54:57 Energy Range (keV): Min 4000 4
2014-09-25 06:54:50 Energy Range (keV): Max 5 with M1; 2.05-20 with M0 or unfocused 18.200
2014-09-25 06:54:34 Energy Range (keV): Min 2.05 4000
2014-09-25 06:54:23 Flux Exponent At 9 keV, 2 x 10^13 At 9 keV, 2 x 10^13 ph/s
2014-09-25 06:54:15 Flux Exponent E(11) ph/sec at 9 keV, ~10E(10) ph/sec at 3 keV, 100 mA At 9 keV, 2 x 10^13
2014-09-25 06:53:06 Mirrors M0: Bent, cylindrical, Pt-coated fused silica; M1: Flat, bent, Ni coated, fused silica M0 mirror: Bent flat single-crystal Si, Rh coated 10-18.2 keV cutoff, vertically collimating M1 mirror: Bent cylindrical single-crystal Si, Rh coated, standard focus
2014-09-25 06:53:01 Optics Details M0 mirror: Bent flat single-crystal Si, Rh coated 10-18.2 keV cutoff, vertically collimating M1 mirror: Bent cylindrical single-crystal Si, Rh coated, standard focus
2014-09-25 06:52:49 Crystal Type Si(111), phi=0 and Si(311) phi=0, Liq N2 cooled; water cooled Si(111), phi=0 and Si(311) phi=0, Liq N2 cooled
2014-09-25 06:52:42 Crystal Type Si(111); water cooled Si(111), phi=0 and Si(311) phi=0, Liq N2 cooled; water cooled
2014-09-25 06:51:54 Optics Details M0 mirror: Bent flat single-crystal Si, Rh coated 10-18.2 keV cutoff, vertically collimating M1 mirror: Bent cylindrical single-crystal Si, Rh coated, standard focus
2014-09-25 06:51:22 Spot Size (mm): Width 0.4 focused 0.4
2014-09-25 06:51:15 Spot Size (mm): Width 3.5 focused; 3 x 20 unfocused (vertical x horizontal) 0.4 focused
2014-09-25 06:50:57 Spot Size (mm): Height 1 0.15
2014-09-25 06:50:44 Source Details Acceptance 2.3mrad focused; 1.0mrad unfocused
2014-02-21 06:29:44 Detector: Model MARMOSAIC 325
2014-02-21 06:29:36 Detector: Type 13-element Ge solid state array; Lytle detector; ion chambers on request Lytle and PIPS detector
2014-02-21 06:29:00 Special Capabilities Facilities for measurements at S K/Cl K edge region for biological samples Facility for advanced X-ray spectroscopy measurements, including XES, RIXS, X-ray Raman and HERFD.
2014-02-21 06:28:20 Experiments FLUORESCENCE XAFS
2014-02-21 06:28:02 Experiments XAS FLUORESCENCE
2014-02-21 06:27:32 Status Details open
2014-02-21 06:27:20 Primary Contact: Email serena@slac.stanford.edu tsuchien@slac.stanford.edu
2014-02-21 06:25:39 Primary Contact: Name Serena DeBeer George Tsu-Chien Weng