| Primary Contact: Name | Andrew McCarthy |
| Secondary Contact: Name | Christoph Mueller-Dieckmann |
| Owner/Operator | EMBL-ESRF Joint Structural Biology Group |
| Status | Operational |
| %Time Available (general use) | 90 |
| Experiments | MAD |
| Experiments Details | A tunable, high intensity, variable focus macromolecular crystallography beamline |
| Special Capabilities | Adjustable Focus Mirrors |
| Special Capabilities Details | Automated high throughput with in situ data collection capability |
| Services | Structure solution |
| Services Details | High speed and quality data collections possible |
| Source Type | INSERTION DEVICE |
| Source Details | 2 1.4m U35 undulators |
| Spot Size (mm): Height | 20 |
| Spot Size (mm): Width | 20 |
| Monochromator | Standard ESRF channel cut |
| Crystal Type | Si(111) |
| Mirrors | Be CRL lenses for vertical focusing and Rh/Pt/Si coated ellipitcal mirror for horizontal focusing |
| Flux(PHOTONS/SEC) | 6 x 10^12 |
| Flux Details | |
| Wavelength Range (Å) | 2.2 - 0.62 |
| Energy Range (keV) | 5.5 - 20.0 |
| Goniometer | Arinax MD2S |
| Sample Handling: Robotics | Other |
| Robotics: More Info | FlexHCD sample changer |
| 1. Detector: Type | PIXEL |
| 1. Detector: Model | DECTRIS PILATUS 6M |
| 1. Detector: Details | PILATUS3 6M |
| 1. Detector: Activated for Service | 2016-08-22 |
| 1. Detector: Removed from Service | |
| 2. Detector: Type | FLUORESCENCE |
| 2. Detector: Model | DECTRIS PILATUS 6M |
| 2. Detector: Details | Ketek AXAS-A |
| 2. Detector: Activated for Service | 2015-03-30 |
| 2. Detector: Removed from Service | |
| Cryo Capability | Oxford |
| Lab Facilities | NORMAL |
| Software: Name | BLISS and MxCuBE3 |
| Software: Classification | Control/collection |
| Hardware | Parallel processing server for data processing/analysis |
| Backup Services | USB3 or remote backup |