Facility Data Managers Login Here

New User? | Forgot Password?

ESRF - ID14-3

Edit this ID14-3 View change history for ID14-3

Beamline Name ID14-3
Primary Contact: NamePetra Pernot
Secondary Contact: NameAdam Round
Owner/OperatorStructural Biology Group
StatusDecommissioned
Status Detailsthe functions of the beamline have been moved to BM29
%Time Available (general use)90
ExperimentsSAXS, SOLUTION STUDIES
Experiments DetailsFrom its inception (October 1997) until it was refurbished in 2002 this was a tuneable beamline (0.93 - 1.44 A). Since January 2003 it has been a fixed wavelength beamline. The beamline has since been re-built from an MX to a bio-SAXS beamline with user operations successfuly started in November 2008. In December 2011 the Bio-SAXS set-up will move to BM29. It will be available for users in June 2012.
Special Capabilitiestemperature control, sample changer
Special Capabilities Detailsroutine data collection on protein in solution in the temperature range from 5 to 50 degrees C
ServicesData collection, Processing
Services Detailsautomatic data collection and semi-automatic processing pipeline
Source Typeinsertion device
Source DetailsU24
Spot Size (mm): Height0.7
Spot Size (mm): Width0.7
Monochromatordiamond (111) in Laue geometry
Optics DetailsGe(220) in Bragg geometry as second crystal
Crystal Typediamond (111) and Ge (220)
Mirrorstorodial focusing mirror
Flux(PHOTONS/SEC)4 x 10^11
Flux Details
Wavelength Range (Å)Fixed: 0.931
Energy Range (keV)Fixed: 13.32
Goniometernone
Goniometer Detailsnone
Sample Handling: Manualsample handing equipment with off-axis visualisation, sample cell is a 2 mm diameter quartz capillary with 10 micron wall thickness
Sample Handling: Roboticssample changer for solutions (EMBL grenoble developement)
Robotics: More Infonew sample changer for solution: storage capacity of 8 samples (0.2 mL PCR tubes) and 3 buffers (1.5 mL tubes)
1. Detector: Type
1. Detector: ModelADSC QUANTUM 4r
1. Detector: Detailsgas filled Vantec 2000 of Bruker
1. Detector: Activated for Serviceinfo currently not available
1. Detector: Removed from Service
2. Detector: TypePIXEL
2. Detector: Model
2. Detector: DetailsPilatus 1M
2. Detector: Activated for Service1997-10-01
2. Detector: Removed from Service2006-01-01
Cryo CapabilityNONE
Lab FacilitiesNORMAL
Software: NameBsxCuBE
Software: Classification
Hardwarehttp://www.esrf.fr/UsersAndScience/Experiments/MX/ID14-3/UserGuide/Computing