Primary Contact: Name | Antoine Royant |
Owner/Operator | French Beamline for Investigation of Proteins (FIP2) |
Status | Operational |
Status Details | Replaces FIP-BM30A |
Experiments | MAD, SAD, MONOCHROMATIC, FLUORESCENCE, HIGH PRESSURE |
Experiments Details | Anomalous difraction experiments, on-line UV-Vis absorption spectroscopy |
Special Capabilities | High Pressure, High Throughput, Humidity Control |
Source Type | BENDING MAGNET |
Source Details | 2-Pole Wiggler in configuration A (2PWA) |
Spot Size (mm): Height | 0.075 to 0.3 |
Spot Size (mm): Width | 0.075 to 0.3 |
Monochromator | double crystal DCM |
Crystal Type | Si (111) |
Mirrors | cylindrical Rh coated |
Flux(PHOTONS/SEC) | |
Flux Details | 7x10^11 ph/s at 12 KeV with 250x250 um beam |
Wavelength Range (Å) | info not available |
Energy Range (keV) | 6 - 21 |
Goniometer | MiniDiff + Kappa |
Sample Handling: Manual | B5 (Spine type) base 21 mm pin lenght |
Sample Handling: Robotics | Flex-HCD |
Robotics: More Info | unipucks |
Detector: Type | PIXEL |
Detector: Model | DECTRIS PILATUS 6M |
Detector: Activated for Service | 2021-06-01 |
Detector: Removed from Service | |
Cryo Capability | Oxford |
Cryo Capability Details | 80 to 400 K |
Lab Facilities | NORMAL |