Primary Contact: Name | Polentarutti Maurizio |
Secondary Contact: Name | Cassetta Alberto |
Owner/Operator | Sincrotrone Trieste and CNR |
Status | Operational |
Experiments | MAD, SAD, MONOCHROMATIC, FLUORESCENCE, HIGH PRESSURE, POWDER, SOLUTION STUDIES |
Experiments Details | The beamline has been designed primarily for macromolecular crystallography.The tunability of the wavelength provides the opportunity to use the Multiwavelength Anomalous Diffraction (MAD) technique for solving the phase problem over a broad range of wavelengths, covering the absorption edges of all the heavy atoms commonly used in protein crystallography. The multipole wiggler spectrum includes high photon flux at low energies, allowing optimisation of the anomalous signal at the Sulphur, Xenon and Calcium edges. Helium purged beampath and helium based cryo-cooling at 100 K are available for data-collection experiments with long wavelengths.The beamline hosts also small molecules crystallography, protein crystallography, powder diffraction, high pressure physics, solid-state experiments. |
Special Capabilities | Absorption Edges, Automated Sample Screening, High Pressure, High Resolution, High Throughput, Long Wavelength |
Special Capabilities Details | The multipole wiggler spectrum includes high photon flux at low energies, allowing optimisation of the anomalous signal for Sulphur, Xenon and Calcium. |
Services | Data collection, On-site cyrstallization facility, Processing, Rapid access |
Source Type | INSERTION DEVICE |
Source Details | NdBFe multipole Wiggler |
Spot Size (mm): Height | 0.200 |
Spot Size (mm): Width | 0.200 |
Monochromator | Double Crystal or white beam |
Crystal Type | Si111 |
Mirrors | Collimating and focusing, Pt-coated |
Flux(PHOTONS/SEC) | 1 x 10^13 |
Flux Details | |
Wavelength Range (Å) | 0.58 - 3.10 |
Energy Range (keV) | 4 - 21.5 |
Goniometer | Huber Kappa Goniometer |
Sample Handling: Robotics | Elettra Sample Changer |
Robotics: More Info | SPINE pins |
1. Detector: Type | CCD |
1. Detector: Model | MAR 165 mm |
1. Detector: Details | Installed on 2theta arm of the Kappa Goniometer |
1. Detector: Activated for Service | info currently not available |
1. Detector: Removed from Service | |
2. Detector: Type | PIXEL |
2. Detector: Model | DECTRIS PILATUS 2M |
2. Detector: Details | The detector operates in single-photon-counting mode and has a large active area of 254x289 mm2, high dynamic range (2,000,000 counts per pixel), high frame rate (30 Hz) and minimal readout time of 3.6 ms. Images can be analyzed via XDS, imosflm and fit2d. |
2. Detector: Activated for Service | 2010-03-01 |
2. Detector: Removed from Service | |
Cryo Capability | Oxford Cryosystems 700 series |
Cryo Capability Details | routinely run at 100K |
Lab Facilities | Cold room, high magnification microscope, microscope for sample mounting |
1. Software: Name | CCP4 suite (iMosflm) |
1. Software: Classification | |
2. Software: Name | XDS |
2. Software: Classification | Processing |
Hardware | Linux PC workstations |
Backup Services | DVD, CD , connection of user's own laptop and USB devices |