Primary Contact: Name | Frank von Delft |
Secondary Contact: Name | Jose Brandao-Neto |
Owner/Operator | Diamond, Harwell Chilton Science Campus, South Oxfordshire, UK |
Status | Operational |
Status Details | I04-1 is the engine for the XChem fragment drug discovery engine. |
%Time Available (general use) | 100 |
Experiments | SAD, MONOCHROMATIC |
Experiments Details | I04-1 uses medium throughput data collection for rapid structure solution using the pan-dataset approach (PanDDAS) - up to 600 measurements per day. https://www.jove.com/t/62414/achieving-efficient-fragment-screening-at-xchem-facility-at-diamond |
Special Capabilities | Automatic Crystal Alignment, High Throughput |
Special Capabilities Details | Single anomalous diffraction (SAD) experiments benefit from energy set near the Bromine edge - most usual elements will give very good anomalous signal for at least great anomalous fourier maps. |
Services | High quality data collection, High speed data collection, Mail-in, Processing, Structure solution |
Spot Size (mm): Height | 0.05 |
Spot Size (mm): Width | 0.05 |
Monochromator | single bounce |
Crystal Type | Si(111) |
Mirrors | toroidal |
Flux(PHOTONS/SEC) | 3.5 x 10^11 |
Flux Details | @ 13530 eV at sample |
Wavelength Range (Å) | Fixed: 0.916 |
Energy Range (keV) | Fixed: 13.53 |
Goniometer | Single axis, extremely accurate air bearing goniometer (a possible addition of a mini-kappa is being explored) |
Sample Handling: Robotics | Diamond BART |
Robotics: More Info | SPINE pins and pucks |
1. Detector: Type | CCD |
1. Detector: Model | MARmosaic 300 |
1. Detector: Activated for Service | 2010-04-10 |
1. Detector: Removed from Service | 2010-11-19 |
2. Detector: Type | PIXEL |
2. Detector: Model | DECTRIS PILATUS 2M |
2. Detector: Details | Pilatus 2M |
2. Detector: Activated for Service | 2010-11-20 |
2. Detector: Removed from Service | 2015-04-14 |
3. Detector: Type | PIXEL |
3. Detector: Model | DECTRIS PILATUS 6M |
3. Detector: Details | 25Hz |
3. Detector: Activated for Service | 2015-04-15 |
3. Detector: Removed from Service | 2021-05-30 |
4. Detector: Type | PIXEL |
4. Detector: Model | Other |
4. Detector: Details | Eiger2 XE 9M 550Hz |
4. Detector: Activated for Service | 2021-06-01 |
4. Detector: Removed from Service | 2021-12-31 |
Cryo Capability | A cryocooling device to keep the sample at 100K |