Facility Data Managers Login Here

New User? | Forgot Password?

APS - 14-ID-B

Edit this 14-ID-B View change history for 14-ID-B

Beamline Name 14-ID-B
Primary Contact: NameVukica Srajer
Secondary Contact: NameRobert Henning
Owner/OperatorBioCARS, CARS, The University of Chicago
StatusOperational
%Time Available (general use)100
ExperimentsMONOCHROMATIC, SAXS, WAXS, TIME RESOLVED LAUE, SOLUTION STUDIES
Experiments DetailsDual in-line undulators (U23 and U27). Supports Laue and monochromatic time-resolved crystallography and time-resolved SAXS/WAXS. Laser facility (ps and ns lasers) is available. Standard macromolecular crystallography is also supported. Biohazardous samples at BSL-2 and BSL-3 levels are supported. Flux listed below is for monochromatic beam at 12keV. In pink beam operation flux is 3x10^10 photons/100ps X-ray pulse in APS hybrid mode. For polychromatic solution scattering accessible q range: 0.02-3.5 A-1.
Special CapabilitiesAdjustable Focus Mirrors, High Intensity, High Resolution, Laue , Biohazards (BSL-2/3), Pulsed Lasers
Special Capabilities DetailsPink beam capability. Time-resolved crystallography and SAXS/WAXS with 100ps time resolution.
ServicesHigh quality data collection, High speed data collection, Rapid access
Source TypeINSERTION DEVICE
Source Detailsdual in-line undulators (U23 and U27)
Spot Size (mm): Height0.015
Spot Size (mm): Width0.020
MonochromatorSi(111) double-crystal
Crystal TypeSi(111)
Mirrorsadjustable focus primary KB mirror pair; Si and Rh stripes; secondary KB mirrors
Flux(PHOTONS/SEC)7 x 10^13
Flux Details@12 KeV
Wavelength Range (Å)0.6 - 1.8
Energy Range (keV)7 - 20
Goniometersingle axis w/ automated crystal centering
Goniometer DetailsTravel ranges: +/-5mm in x and y; +/- 60mm in z
Detector: TypeCCD
Detector: ModelRayonix MX340-HS
Detector: DetailsReadout rate: 10-100Hz
Detector: Activated for Service2014-02-01
Detector: Removed from Service
Cryo CapabilityOxford Cryojet
Lab FacilitiesLaser Lab (ps and ns lasers), Biosafety Cabinet, Cold Room
Software: NameBeamline control: EPICS; Data acquisition: BioCARS-developed GUI; Data processing: HKL2000, Denzo/HKL, Mosflm, Precognition (Laue processing), CCP4, Coot; for time-resolved solution scattering (SAXS/WAXS) BioCARS-developed data reduction package
Software: ClassificationControl/collection, Processing, Phasing/refinement, Modeling/graphics
Backup ServicesUSB drives and Globus services