Primary Contact: Name | Matthew Latimer |
Secondary Contact: Name | Erik Nelson |
Owner/Operator | SSRL-SMB-XAS |
Status | Operational |
Status Details | Open |
%Time Available (general use) | 100 |
Experiments | XAFS, EXAFS |
Experiments Details | Beam line 4-3 is a wiggler side-station beam line for X-ray absorption spectroscopy and EXAFS measurements. This station enables tender X-ray measurement (S K-edge and up) and is setup with a He flight path from the beam line optics to the sample. The beam is collimated and unfocused to allow for high energy-resolution measurements on homogenous samples. 4-3 is equipped with a 4-element solid-state Si drift detector in addition to ionization chambers, PIPS and Lytle detectors. A liquid He cryostream allows for low-temperature measurements. A motorized sample holder stage allows sequential measurement on multiple samples without sample changeover delays. |
Source Type | INSERTION DEVICE |
Source Details | 20-pole, 2-Tesla wiggler |
Spot Size (mm): Width | 20 |
Monochromator | Liquid nitrogen-cooled double crystal, non-fixed exit slit. |
Crystal Type | Si(111) |
Mirrors | Vertically collimating, flat, bent, 1.0 m, Si, Ni-coated, cutoff variable 4.5 to 14.5 keV |
Flux(PHOTONS/SEC) | 1 x 10^12 |
Flux Details | |
Wavelength Range (Å) | info not available |
Energy Range (keV) | 2.4 - 14.5 |
Detector: Type | Ionization chambers, PIPS, Lytle, Hitachi HTA 4-element solid-state Si drift detector, electron-yield detector. |
Detector: Model | |
Detector: Activated for Service | info currently not available |
Detector: Removed from Service | |
Lab Facilities | Sample preparation laboratory, cold room, glove boxes |
Software: Name | XASCollect, Xasscan, EXAFSPAK, SixPAK |
Software: Classification | Control/data collection, Processing, Analysis |