| Primary Contact: Name | Matthew Latimer |
| Secondary Contact: Name | Erik Nelson |
| Owner/Operator | SSRL-SMB-XAS |
| Status | Operational |
| Status Details | Open |
| %Time Available (general use) | 100 |
| Experiments | XAFS, EXAFS |
| Experiments Details | Beam line 4-3 is a wiggler side-station beam line for X-ray absorption spectroscopy and EXAFS measurements. This station enables tender X-ray measurement (S K-edge and up) and is setup with a He flight path from the beam line optics to the sample. The beam is collimated and unfocused to allow for high energy-resolution measurements on homogenous samples. 4-3 is equipped with a 4-element solid-state Si drift detector in addition to ionization chambers, PIPS and Lytle detectors. A liquid He cryostream allows for low-temperature measurements. A motorized sample holder stage allows sequential measurement on multiple samples without sample changeover delays. |
| Source Type | INSERTION DEVICE |
| Source Details | 20-pole, 2-Tesla wiggler |
| Spot Size (mm): Width | 20 |
| Monochromator | Liquid nitrogen-cooled double crystal, non-fixed exit slit. |
| Crystal Type | Si(111) |
| Mirrors | Vertically collimating, flat, bent, 1.0 m, Si, Ni-coated, cutoff variable 4.5 to 14.5 keV |
| Flux(PHOTONS/SEC) | 1 x 10^12 |
| Flux Details | |
| Wavelength Range (Å) | info not available |
| Energy Range (keV) | 2.4 - 14.5 |
| Detector: Type | Ionization chambers, PIPS, Lytle, Hitachi HTA 4-element solid-state Si drift detector, electron-yield detector. |
| Detector: Model | |
| Detector: Activated for Service | info currently not available |
| Detector: Removed from Service | |
| Lab Facilities | Sample preparation laboratory, cold room, glove boxes |
| Software: Name | XASCollect, Xasscan, EXAFSPAK, SixPAK |
| Software: Classification | Control/data collection, Processing, Analysis |