| Primary Contact: Name | Aina Cohen |
| Secondary Contact: Name | Mike Soltis |
| Owner/Operator | SSRL-SMB-MC |
| Status | Operational |
| %Time Available (general use) | 100 |
| Experiments | MAD, SAD, MONOCHROMATIC |
| Experiments Details | Full Remote Access |
| Special Capabilities | Absorption Edges, Adjustable Focus Mirrors, Automatic Crystal Alignment, Automated Sample Screening, High Intensity, High Resolution, High Throughput, Micro-diffraction, Multilayer Mode |
| Special Capabilities Details | Full remote access |
| Services | High quality data collection, High speed data collection, Processing, Rapid access, Structure refinement, Structure solution, On-site crystallization facility |
| Services Details | In-vacuum mini-gap undulator |
| Source Type | INSERTION DEVICE |
| Spot Size (mm): Height | 0.005 - 0.2 |
| Spot Size (mm): Width | 0.04 - 0.15 |
| Monochromator | Liquid Nireogen Double Crystal |
| Crystal Type | Si (111) |
| Mirrors | KB |
| Flux(PHOTONS/SEC) | 4 x 10^12 |
| Flux Details | @ 0.98 A |
| Wavelength Range (Å) | 0.69 - 2.0 |
| Energy Range (keV) | 6.2 - 18 |
| Goniometer | SSRL single-axis microdiffractometer |
| Goniometer Details | Air bearing; coaxial cold stream; 1.5 micron SOC |
| Sample Handling: Manual | no restrictions |
| Sample Handling: Robotics | (SAM) Stanford Automated Mounting System |
| Robotics: More Info | https://smb.slac.stanford.edu/facilities/hardware/SAM/ |
| 1. Detector: Type | PIXEL |
| 1. Detector: Model | DECTRIS EIGER 2XE 16M |
| 1. Detector: Details | 100 Hz readout. Up to 750 Hz readout for selected area. Dead Time = 2 us. |
| 1. Detector: Activated for Service | 2022-10-01 |
| 1. Detector: Removed from Service | |
| 1. Removed from Service: Details | Currently in use |
| 2. Detector: Type | PIXEL |
| 2. Detector: Model | DECTRIS EIGER 16M |
| 2. Detector: Activated for Service | 2020-10-12 |
| 2. Detector: Removed from Service | 2022-10-01 |
| 3. Detector: Type | PIXEL |
| 3. Detector: Model | DECTRIS PILATUS 6M |
| 3. Detector: Activated for Service | 2012-10-12 |
| 3. Detector: Removed from Service | 2020-10-12 |
| Cryo Capability | Oxford Cryojet |
| Lab Facilities | Normal |
| Software: Name | See: http://smb.slac.stanford.edu/facilities/software/ |
| Software: Classification | Control/collection, Processing, Phasing/refinement, Modeling/graphics |
| Hardware | http://smb.slac.stanford.edu/facilities/computing/ |
| Backup Services | http://smb.slac.stanford.edu/facilities/computing/backup.html |