| Primary Contact: Name | Howard Robinson |
| Secondary Contact: Name | Wuxian Shi |
| Owner/Operator | PXRR Brookhaven Biology and Case Proteomics Center-Case Western Reserve |
| Status | This beamline was shut down September 30, 2014 |
| Status Details | This beamline was shut down on September 30, 2014 |
| %Time Available (general use) | 25 |
| Experiments | MAD, SAD |
| Source Type | insertion device |
| Source Details | 54 pol in vacuum undulator to 3.3 mm gap |
| Spot Size (mm): Height | 0.15 |
| Spot Size (mm): Width | 0.15on aperture 65 mm upstream of crystal |
| Monochromator | Rosenbaum-Rock double crystal sagittal focusing monochromator |
| Mirrors | vertical focusing mirror |
| Flux(PHOTONS/SEC) | 3.5 x 10^11 |
| Flux Details | at 1.1 A through 120 by 120 aperture with ring at 300 ma |
| Wavelength Range (Å) | 0.83 - 1.8 |
| Energy Range (keV) | 7 - 15 |
| Goniometer | air bearing omega with mini-kappa (to 50 degrees) and motorized crystal positioning |
| Sample Handling: Robotics | ALS type automounter |
| Robotics: More Info | www.px.nsls.bnl.gov/robot/robot.html |
| Detector: Type | CCD |
| Detector: Model | ADSC QUANTUM 315 |
| Detector: Activated for Service | info currently not available |
| Detector: Removed from Service | |
| Cryo Capability | Oxford Cryosystems Cryostream 700 |
| Lab Facilities | no |
| Software: Name | CBASS (with EPICS), HKL2000, MOSFLM |
| Software: Classification | |
| Hardware | Linux PC's |
| Backup Services | DVD burner robot, firewire disks |