Facility Data Managers Login Here

New User? | Forgot Password?

APS - 18-ID

Edit this 18-ID View change history for 18-ID

Beamline Name 18-ID
Primary Contact: NameThomas Irving
Secondary Contact: NameSrinivas ChakravarthyLiang Guo
Owner/OperatorBioCAT
Statusoperational
%Time Available (general use)80
ExperimentsSAXS
Experiments DetailsThe beamline is optimized for non-crystalline diffraction and scattering applications such as static and time-resolved fiber diffraction from muscle, connective tissues and oriented macromolecular assemblies, and conventional and time-resolved scattering from macromolecules in solution.
Source Typeinsertion device
Source Details3.3 undulator
Spot Size (mm): Height0.15
Spot Size (mm): Width0.05
Optics Detailssagital focussing second monochromator crystal vertical focussing Miror
Crystal TypeSi (111)
Mirrorselipsoidal
Flux(PHOTONS/SEC)2 x 10^13
Flux Details@10keV
Wavelength Range (Å)0.035 - 0.3
Energy Range (keV)4 - 35
Goniometer 0.3m - 3.5 m SAXS instrument 0.1-0.4m fiber WAXS instrument
1. Detector: TypeCCD
1. Detector: Model
1. Detector: Activated for Serviceinfo currently not available
1. Detector: Removed from Service
2. Detector: TypeFLUORESCENCE
2. Detector: ModelOther
2. Detector: DetailsVortex detector
2. Detector: Activated for Service2011-01-03
2. Detector: Removed from Service
3. Detector: TypeCCD
3. Detector: ModelOther
3. Detector: DetailsAviex 16080
3. Detector: Activated for Service2008-01-02
3. Detector: Removed from Service
4. Detector: TypePIXEL
4. Detector: ModelOther
4. Detector: DetailsPilatus 100k
4. Detector: Activated for Service2010-11-01
4. Detector: Removed from Service
5. Detector: TypePIXEL
5. Detector: ModelPILATUS IM
5. Detector: DetailsPilatus 3
5. Detector: Activated for Service2014-09-01
5. Detector: Removed from Service
Lab FacilitiesComprehensive biochemical laboratory