Primary Contact: Name | Lisa Keefe |
Owner/Operator | IMCA-CAT |
Status | operational |
%Time Available (general use) | 25 |
Experiments | MAD, MONOCHROMATIC |
Source Type | insertion device |
Source Details | 2.7 undulator (upstream) - 2.3 undulator (downstream) |
Spot Size (mm): Height | 0.5 |
Spot Size (mm): Width | 0.5 mm maximum, 0.05mm x 0.05 mm minimum (unfocused; defined by slits) |
Flux(PHOTONS/SEC) | 7 x 10^11 |
Flux Details | 200 micron aperture @12keV |
Wavelength Range (Å) | 2.0 - 0.7 |
Energy Range (keV) | 6 - 17 |
Goniometer | Equipment in the experiment end station includes CrystalLogic goniostat with omega and theta rotations, Rigaku/MSC ACTOR robot for automated crystal mounting, positioning, and retrieval, and Rontec X |
Sample Handling: Robotics | Rigaku ACTOR |
1. Detector: Type | CCD |
1. Detector: Model | ADSC QUANTUM 210 |
1. Detector: Details | The ADSC Q210 is availalbe as backup |
1. Detector: Activated for Service | info currently not available |
1. Detector: Removed from Service | |
2. Detector: Type | PIXEL |
2. Detector: Model | DECTRIS PILATUS 6M |
2. Detector: Activated for Service | 2012-04-01 |
2. Detector: Removed from Service | |
Cryo Capability | Oxford CryoJet |
Lab Facilities | Biochemistry laboratory and cold room for sample preparation. |
Software: Name | Data collection: JDirector. Data processing: HKL-2000, d*TREK, MOSFLM, XDS. Beamline control: MX. |
Software: Classification | |
Hardware | Linux workstations |
Backup Services | CD-R, DVD-R, external Hi-speed USB, and FireWire (IEEE-1394a) hard drives, user's laptop computer via Gigabit Ethernet |