Primary Contact: Name | Sandor Brockhauser |
Secondary Contact: Name | Andrew McCarthy |
Owner/Operator | Joint Structural Biology Group |
Status | Decommissioned |
Experiments | MAD, SAD, FLUORESCENCE |
Special Capabilities | High Intensity |
Special Capabilities Details | Pipeline dedicated to high throughput structural biology |
Services | Data collection |
Services Details | High-quality and high-speed data collections |
Source Type | insertion device |
Source Details | U35, U24 and U23 undulators |
Spot Size (mm): Height | 0.09 |
Spot Size (mm): Width | 0.250 |
Monochromator | channel cut ESRF monochromator |
Crystal Type | Si(111) |
Mirrors | Toroidal focusing mirror |
Flux(PHOTONS/SEC) | |
Flux Details | ~5 x e12 at 13.2 keV |
Wavelength Range (Å) | 0.939 - 1.3 |
Energy Range (keV) | 9.6 - 14.5 |
Goniometer | Maatel Mini-diffractometer (MD2M) |
Goniometer Details | A minikappa goniometer head (MK3) for optimal crystal reorientation |
Sample Handling: Robotics | EMBL Grenoble Sample Changer (SC3) |
Robotics: More Info | The Sample Changer has been designed for use with SPINE standard sample holders and vials, which are loaded into the SC in baskets specifically designed for the SCs. The goniometer head that must be used in conjunction with the SCs is equipped with a SmartMagnet. This detects whether a sample is mounted on the goniometer head or not. Each indivudual SmartMagnet is tuned for operation on a particular beam-line and a goniometer head must NOT be taken from one beam-line for use on another. |
Cryo Capability | Oxford Cryostream 700 series |