| Primary Contact: Name | Ana Gonzalez |
| Secondary Contact: Name | Thomas Ursby |
| Owner/Operator | MAX IV |
| Status | Operational |
| %Time Available (general use) | 50% |
| Experiments | MAD, SAD, MONOCHROMATIC |
| Experiments Details | Remote access, HT (fully automated) data collection, in situ crystal screening, room temperature data collection, SSX fixed-target |
| Special Capabilities | Absorption Edges, Adjustable Focus Mirrors, Automatic Crystal Alignment, Automated Sample Screening, High Intensity, High Resolution, Humidity Control, Microfocus, Serial Crystallography |
| Services | Mail-in, On-site cyrstallization facility, Processing, Rapid access |
| Services Details | FragMAX platform for fragment-based drug discovery offers the above services. Mail-in is also available for proprietary users |
| Source Type | INSERTION DEVICE |
| Source Details | in vacuum undulator, 18mm period length |
| Spot Size (mm): Height | 0.005 |
| Spot Size (mm): Width | 0.020 |
| Monochromator | Double crystal monochromator |
| Optics Details | Horizontally deflecting |
| Crystal Type | Si111 |
| Mirrors | KB-Si mirror system |
| Flux(PHOTONS/SEC) | 1 x 10^13 |
| Flux Details | 0.1 x 0.1 mrad^2 hxv divergence |
| Wavelength Range (Å) | 0.5 - 2.5 |
| Energy Range (keV) | 5 - 25 |
| Goniometer | MD3-Down |
| Goniometer Details | MiniKappa head (default), plate adaptor |
| Sample Handling: Manual | SPINE sample holders are preferred |
| Sample Handling: Robotics | ISARA |
| Robotics: More Info | UNI-pucks |
| 1. Detector: Type | PIXEL |
| 1. Detector: Model | DECTRIS EIGER 16M |
| 1. Detector: Activated for Service | 2016-10-27 |
| 1. Detector: Removed from Service | 2024-03-25 |
| 2. Detector: Type | PIXEL |
| 2. Detector: Model | DECTRIS EIGER2 XE 16 M CdTe |
| 2. Detector: Details | 311x328 mm^2 active area, 75x75 micrometer ^2 pixel size |
| 2. Detector: Activated for Service | 2024-03-25 |
| 2. Detector: Removed from Service | |
| Cryo Capability | Oxford Cryojet |
| Cryo Capability Details | 95-300 K |
| Lab Facilities | Sample preparation lab, bio-safety level 1 lab, cold room |
| Backup Services | GLOBUS, sftp |
| Serial Crystallography Details | Fixed target |
| Sample Dehydration Prevention Method | HCLab humidifier |
| Fixed Target Mechanism to Hold Sample | Chips |
| Fixed Target Sample Support Base | Goniometer |