Primary Contact: Name | Takaaki HIKIMA |
Owner/Operator | RIKEN Harima Institute |
Status | operational |
%Time Available (general use) | 20 |
Experiments | MAD, SAXS |
Experiments Details | small-angle x-ray scattering (SAXS) and protein crystallography (PX). Both experiments can be independently carried out. Only SAXS station is opening for the public use |
Special Capabilities | High pressure |
Special Capabilities Details | SAXS-station; High-resolution small-angle scattering camera, Stopped-flow mixer, Temperature jump cell, High-pressure cell for protein solutions. |
Source Type | insertion device |
Source Details | in vacuum undulator |
Spot Size (mm): Height | 0.4 mm (horizontal) |
Spot Size (mm): Width | 0.2 mm (vertical) (SAXS), 0.3 mm (horizontal) x 0.2 mm (vertical) (PX) |
Monochromator | Double-crystal monochromator (synthetic diamond (111) in Laue-Bragg geometry) |
Mirrors | K-B mirrors (vertical and horizontal) |
Flux(PHOTONS/SEC) | 3 x 10^^11 photons/sec (SAXS), 2 x 10^11 photons/sec (PX) |
Flux Details | |
Wavelength Range (Å) | 0.886 - 1.653 |
Energy Range (keV) | 7.5 - 14.0 |
Goniometer | Two individual single rotation axis (horizontal and vertical) goniometers are available |
1. Detector: Type | IMAGE PLATE |
1. Detector: Model | RIGAKU R-AXIS V |
1. Detector: Activated for Service | info currently not available |
1. Detector: Removed from Service | |
2. Detector: Type | CCD |
2. Detector: Model | RIGAKU JUPITER 210 |
2. Detector: Activated for Service | info currently not available |
2. Detector: Removed from Service | |
Cryo Capability | cryostream coolers -nitrogen gas |
Lab Facilities | Biology and chemistry preparation laboratories, cold room |