| Beamline Website |
http://smb.slac.stanford.edu/ |
| Owner/Operator |
SSRL-SMB-MC |
| Operational (general use) |
yes |
| %Time available (general use) |
100 |
| Beamline Schedule |
http://www-ssrl.slac.stanford.edu/protein_crystallography_schedule.html |
| Experiments |
MAD, SAD, automated sample screening |
| Source |
16 pole wiggler |
| Ring Current(mA) |
100 |
| Average Beam Lifetime |
70 hrs (90 mA average over 8 hr fill period) |
| Spot Size |
0.6 x 0.21 mm |
| Optics |
Flat collimating mirror, double crystal monochromator, toroid focusing mirror |
| Flux |
2.5x10E(11) ph/s |
| Flux Details |
Flux measured through a 0.2 x 0.2 mm2 aperture at 12658 eV |
| Wavelength Min (Å) |
0.85 |
| Wavelength Max (Å) |
2.07 |
| Monochromator |
Double crystal monochromator |
| Crystal type |
Si(111). A second set of Si(220)
crystals is also available.
|
| Mirrors |
Rh coated |
| Mirror Cutoff |
Variable |
| Special Capabilities |
Automated sample mounting |
| Other Instrumentation |
See:
http://smb.slac.stanford.edu/facilities/hardware/
|
| Goniometer |
Huber Kappa-geometry goniometer |
| Pin Length |
Manual sample mounting: no restrictions; Automated sample mounting: 14-18 mm |
| Restrictions |
omega 75 degs., kappa 20 degs., phi no limits |
| xtal-detector Distance |
95-650 mm |
| Detector |
MarUSA MarMosaic -325 CCD |
| Detector Details |
Inter-image total dead time: 5 s |
| 2theta range |
vertical offset 200 mm ;
horizontal offset 3 mm
|
| Cryo Capability |
Oxford Cryojet |
| Lab Facilities |
Biochemical lab with cold room |
| Biohazard |
P2 Biohazards |
| Software |
See: http://smb.slac.stanford.edu/facilities/software/ |
| Hardware |
See: http://smb.slac.stanford.edu/facilities/computing/ |
| Backup Devices |
See: http://smb.slac.stanford.edu/facilities/computing/backup.html |
|
|