Hosted By the RCSB PDB
BIOSYNC   Structural Biology Synchrotron Users Organization
   Home  |  US Synchrotrons  |  International Synchrotrons  |  Structural Genomics

 

SSRL Beamlines: BL1-5 , BL4-2 , BL6-2 , BL7-1 , BL7-3 , BL9-1 , BL9-2 , BL9-3 , BL11-1 , BL11-3 , BL12-2

Beamline Name BL9-1
Beamline Website http://smb.slac.stanford.edu/
Owner/Operator SSRL-SMB-MC
Operational (general use) yes
%Time available (general use) 100
Beamline Schedule http://www-ssrl.slac.stanford.edu/protein_crystallography_schedule.html
Experiments MAD (Se, Kr, Br, Pb), SAD, automated sample screening
Source 16 pole wiggler
Ring Current(mA) 100
Average Beam Lifetime 70 hrs (90 mA average over 8 hr fill period)
Spot Size 2.2 mm x 0.23 mm
Optics Vertical focusing mirror; single crystal Si(311) bent monochromator (horizontal focusing)
Flux 4.1x10E(10) ph/s
Flux Details Flux measured through a 0.2 x 0.2 mm2 aperture at 12658 eV and 92 mA current
Wavelength Min (Å) 0.79
Wavelength Max (Å) 0.99
Monochromator Side-scattering cuberoot I-beam bent single crystal; asymetric cut 12.2 degs.
Crystal type Si(311)
Mirrors Flat mirror, Rh coated
Mirror Cutoff 16keV
Special Capabilities Automated sample mounting
Other Instrumentation See: http://smb.slac.stanford.edu/facilities/hardware/
Goniometer Huber Kappa-geometry goniometer
Pin Length Manual sample mounting:no restrictions; Automated sample mounting: 14 - 18 mm
Restrictions omega 90 degs., kappa 35 degs., phi no limits
xtal-detector Distance 95-650 mm
Detector ADSC Quantum-315R CCD
Detector Details Inter-frame total dead time: 3s.
2theta range vertical offset 150mm; horizontal offset 150mm
Cryo Capability Oxford Cryojet
Lab Facilities Biochemical lab with cold room
Biohazard P2 Biohazards
Software See: http://smb.slac.stanford.edu/facilities/software/
Hardware See: http://smb.slac.stanford.edu/facilities/computing/
Backup Devices See: http://smb.slac.stanford.edu/facilities/computing/backup.html
spacer spacer


Update this site - synchrotron administrators only