SSRL Beamlines: BL1-5 , BL4-2 , BL6-2 , BL7-1 , BL7-3 , BL9-1 , BL9-2 , BL9-3 , BL11-1 , BL11-3 , BL12-2

Beamline Name BL11-1
Beamline Website http://smb.slac.stanford.edu/
Owner/Operator Scripps Research Institute (TSRI), Stanford University and SSRL-SMB-MC
Operational (general use) yes
%Time available (general use) 33%
Beamline Schedule http://www-ssrl.slac.stanford.edu/protein_crystallography_schedule.html
Experiments MAD (restricted energy range), SAD, Automated sample screening
Source 26 pole wiggler
Ring Current(mA) 100
Average Beam Lifetime 70 hrs (90 mA average over 8 hr fill period)
Spot Size 1.2 mm x 0.3 mm
Optics Flat mirror (vertical focusing); single crystal Si(111) bent monochromator (ho rizontal focusing)
Flux 3.5 x 10E(11)
Flux Details Flux measured through a 0.2 x 0.2 mm2 aperture at 12658 eV and 94 mA current
Wavelength Min (Å) 0.82
Wavelength Max (Å) 1.17
Monochromator Side scattering bent cube-root I-beam single crystal; asymmetric cut 4.965 degs
Crystal type Si(111)
Mirrors Rh coated flat mirror
Mirror Cutoff ?
Special Capabilities Automated sample mounting
Other Instrumentation See: http://smb.slac.stanford.edu/facilities/hardware/
Goniometer Huber Kappa-geometry goniometer
Pin Length Manual sample mounting: no restrictions; Automated sample mounting: 14-18 mm
Restrictions omega 65 degs., kappa 35 degs., phi no limits
xtal-detector Distance 90-650 mm
Detector MarUSA MarMosaic -325 CCD
Detector Details Inter-image dead time: 5 s.
2theta range vertical offset: 175 mm horizontal offset: 180 mm
Cryo Capability Oxford Cryojet
Lab Facilities Beamline cold room Biochemical lab with cold room
Biohazard P2 Biohazards
Software See: http://smb.slac.stanford.edu/facilities/software/
Hardware See: http://smb.slac.stanford.edu/facilities/computing/
Backup Devices See: http://smb.slac.stanford.edu/facilities/computing/backup.html
spacer spacer