ELETTRA Beamlines: 5.2R

Beamline Name 5.2R
Beamline Website http://www.elettra.trieste.it/experiments/beamlines/xrd1/index.html
Owner/Operator ?
Operational (general use) yes
%Time available (general use) ?
Beamline Schedule http://www.elettra.trieste.it/accelerator/shifts.php
Experiments SAD - MAD - Powder - High Pressure
Source NdBFe multipole wiggler
Ring Current(mA) ?
Average Beam Lifetime ?
Spot Size 1.2 x 1.2 mm2
Optics Double crystal monochromator with toroidal mirrors
Flux 1012-1013 ph/sec
Flux Details ?
Wavelength Min (Å) 0.50
Wavelength Max (Å) 3.10
Monochromator Double Crystal
Crystal type Si111 - Si220
Mirrors Three-segment Pt-coated toroidal
Mirror Cutoff ?
Special Capabilities ?
Other Instrumentation x-ray detector for acquisition of fluorescence spectra and absorption edge determination for MAD experiments
Goniometer Single axis goniostat (horizontal)
Pin Length ?
Restrictions ?
xtal-detector Distance ?
Detector Mar 345 Image Plate, Mar 165 CCD
Detector Details ?
2theta range ?
Cryo Capability 700 series Oxford Cryosystems
Lab Facilities Cold room, high magnification microscope, microscope for sample mounting
Biohazard ?
Software HKL suite (Xdisp, Denzo and Scalepack), Mosflm and XDS
Hardware Linux PC workstations, SG Indy and O2 workstations, 4-processor Compaq Alpha SE server.
Backup Devices 4mm DAT, DVD, CD and connection of user's own laptop
spacer spacer